ProductIntroduce: AED7000 AUTOMATIC EXTERNAL DEFIBRILLATOR
l Three-step defibrillation process
l Two-button operation
l Extensive voice and visual prompts for the operator
l Biphasic energy output
l Lock-out protection to prevent inadvertent defibrillation
Specification
l Defibrillator
Category
|
Specification
|
Output
|
Biphasic truncated exponential
|
Energy sequence
|
150,150,200J
|
Charge time
|
8sec.to 150J, 10sec to 200J(New Battery)
|
Analysis time
|
9sec
|
The Maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery
|
Less than 30 seconds
|
The Maximum time from the initiation of rhythm analysis to readiness for discharge after 6 shocks
|
Less than 35seconds
|
The Maximum time from initially switching power on to readiness for discharge
|
Less than 25 seconds
|
Audible prompts
|
22 audible prompts
|
Visual prompts
|
LED prompts
|
Controls
|
Two Buttons-On/Off, Shock
|
Output energy accuracy
|
±15%into any impedance from 25 to 175Ω
|
The maximum voltage
|
1200±50V
|
Output disabled when the patient impedance is outside limits
|
20Ω to 200Ω
|
l Battery
Part No
|
CR123A-4*2
|
Non-Rechargeable
|
12V D.C.2.8Ah
|
Capacity
|
100 discharges at 200 joules or 120 discharges at 150 joules
|
l Physical
Dimension
|
303*216*89mm
|
Weight
|
2.0kg
|
Operating temperature
|
0℃ to 40℃
|
Operating humidity
|
Relative humidity between 30% and 95%(non-condensing _
|
Storage temperature(without battery)
|
-20℃ to 55℃
|
Storage humidity(without battery)
|
Up to 95%(non-condensing )
|
|