Opitcal Performance
Optical System: Littow type, Single beam, Grating system 1200 1/mm
Spectral Slitwidth: 2nm
Wavelength Range: 190-1100nm
Accuracy:±0.5nm
Repeatability:0.3nm
Stray Light: ≤0.1%T@220&340nm
Scan speed: fast, medium, slow
|
Photometric Performance
Photometric Mode: Transmittance
Absorbance,
Concentration
Photometric Range: 0-125.0%T,-0.097-3.000A,0-9999C
Accuracy: ±0.3%T(0-100%T)
±0.002(0-0.5Abs)
±0.004(0.5-1Abs)
Repeatability: 0.2%T(0-100%T)
0.001(0-0.5Abs)
0.002(0.5-1Abs)
Base Line: ±0.003A
±Drift: ±0.002A/h(at 500nm after warm up 1 hour)
±Noise:±0.002A
|
Main Functions
100%T: Auto
0%T: Auto
Auto self-test wavelength, light source and filter
Date Output: Analog RS-232C
PC Controlled: Required Win-spec.3.33EN Software
Win-spec 3.33EN:
Spectrum scanning
Abs/T%/Conc Measurement
Standart curve
Time scanning
|
Specification
Display: 2-line, 20-character backlighted LCD
Sample Compartment: 100mm light beam for test Built in 4 position Cell Holder: 10mm
Power Requirement: 110/220VAC, 50/60Hz, ±10%Built in 4 position Cell Holder: 10mm
Standard setup of Cells: 10mmGlass cell 4pcs and Quartz cell 2pcs Each
|